The zero point on an electronic apex locator
Which anatomical region does the 0.0 / APEX mark on an electronic apex locator correspond to? How does it relate to the minor apical diameter, the major apical foramen and Black's space? We see different explanations in the literature. Which should we use as a clinical reference?
Translated version of the Turkish original
The APEX mark on the Root ZX II and the 0.0 / APEX mark on the Woodpecker Woodpex V and Ai‑Pex target the major apical foramen. On the Root ZX II, the 0.5 mark is a reference for the apical constriction region. On the Woodpecker Smart A, intermediate scale marks such as 0.1, 0.2 and 0.3 indicate that the file tip is approaching the foramen (1). www.morita.com
The distance from the apical constriction to the major foramen varies with canal anatomy. In a micro-CT study of molar canals, the mean anatomical distance was 0.2 mm, with a measured range of 0–0.6 mm. The 0.2 mm in the study is an anatomical measurement. The 0.2 on the Woodpecker display is a scale mark (2). pubmed.ncbi.nlm.nih.gov
Black's space is classically described as the funnel-shaped region that widens from the minor apical diameter towards the major foramen. The major foramen targeted by the APEX mark on these devices is the end of Black's space that opens into the periodontal tissues (3).