Latest article: Choosing the appropriate file taper

Reciprocation angles

When should we choose reciprocation angles of 30°–150°, 30°–170° or the reverse? What determines the direction of the larger angle?

Translated version of the Turkish original

The larger angle should be in the direction in which the file cuts dentine. If the file cuts anticlockwise (CCW), the larger angle is CCW. If it cuts clockwise (CW), the larger angle is CW. The smaller angle in the opposite direction helps disengage the file from the dentine.

When the cutting and release angles are 150° and 30° respectively, the file advances a net 120° in the cutting direction per cycle. At 170° and 30°, the net advance is 140°. The sizes of both angles can affect cutting efficiency and the torque generated.

In a laboratory study using WaveOne, a narrower reciprocation range increased cyclic fatigue resistance and prolonged preparation time (1). In our laboratory study on Reciproc R25, 150° CCW / 30° CW also produced the longest time to fracture among the settings tested (2). These findings are specific to the file and test conditions used.

Author of the original answer: Doç. Dr. İsmail Davut Çapar

Medical review of the Turkish original:

Turkish answer published on this website:

Translation published:

The review date refers to the Turkish original; it does not indicate a separate medical review of the translation.

References